Opens in a new tab
  • BIT-L01S

    CCIT – Bacterial Ingress Tester

    External air source

    The BIT-L01S container closure integrity tester (Dye Penetration/Microbial Challenge Method) is a specialized device for ensuring the integrity of sterile pharmaceutical packaging. It provides a comprehensive solution by integrating pressure control, microbial solution management, and protection, streamlining the entire process of microbial challenge packaging integrity testing.

    01
    Self-Diagnostic Fuction
    02
    Adjustable Parameters
    03
    Permission Management

    Model

    Features

    • Supports dye penetration and microbial challenge methods for sterile packaging integrity testing
    • Storage of multiple testing protocols, simplifying parameter configuration
    • Suitable for various packaging types, including powder injections and large-volume injectables
    • Custom-designed immersion racks ensure complete contact between samples and microbial solution
    • Automatically saves test data, which can be reviewed but not modified
    • Real-time data display via a color touchscreen
    • Switchable negative and positive pressure settings for more rigorous and consistent testing
    • Main unit and test tank are separate, with quick-connect pipelines to ensure safety and prevent leakage
    • Stainless steel test tank can endure 125°C steam sterilization, with flange-sealed openings to prevent leakage,
    • and an IP69 waterproof rating
    • All pressure adjustments and microbial solution management are conducted in a fully sealed environment
    • Traceable operations with exportable, unmodifiable records

    Specifications

    Main Unit Dimensions: 415 mm×430 mm×240 mmWeight: Approx. 18 kg
    Test Tank Capacity: 10L, 20L, 30L (customizablePower: 220 V, ≤500 W
    Formula Storage: 100Alarms: 1. High or low tank pressure; 2. Self-check failure
    Pressure Range: -90~+400 KPaPressure Accuracy: ±0.5 KPa
    Automatic Bacterial Solution Filling: Speed-adjustable peristaltic pump(optional)Sample Loading Capacity: Depends on tank size and sample rack (customizable)

    Compliance

    • Complies with GAMP5 validation standards
    • Adheres to NMPA CDE guidelines for the integrity of chemical injectable packaging
    • Fulfills USP43 1207.2 requirements for packaging integrity testing
    • Aligns with T/ZJPA 001-2023 standards for microbial challenge immersion tests in injectable packaging

    Applications

    • Challenge tests for sterile packaging during transportation and storage
    • Pharmaceutical stability research and validation
    • Studies on the correlation between physical pore size and microbial penetration
    • Validation of physical inspection methods
    • Verification of crimping and sealing processes

    Pharmaceutical liquid preparation system

    Radiopharmaceutical production

    Disposable filtration system PUPSIT

    Send Inquiry