Applied in Biosafety Level 3(BSL-3) and above laboratories, specifically designed for the rearing, experimentation, and other operations of laboratory ferrets, fulfilling both biosafety and animal experimentation requirements.
Multiple housing densities available with single or double chamber structures;independent control ensures no mutual interference.Parallel redundant exhaust units automatically monitor pressure loss to guarantee a negative pressure environment,featuring online fan replacement capability.
Configured with multi-stage exhaust filtration (mesh,primary,and high-efficiency filters)to provide a safe and effective environmental guarantee.
Terminal exhaust equipped with H14 HEPA filters(99.995%MPPS efficiency per EN1822), featuring automatic scan leak detection,efficiency leak testing, in-situ disinfection, and Bag-In-Bag-Out(BIBO) replacement.
Front operating side features viewing windows with gloves and multi-point mechanical compression sealing for stable biosafety assurance.
The front face adopts an ergonomic tilted design for ease of operation.
Rigid chamber structure with rounded corners, free of burrs, sharp edges, and cleaning dead zones; chamber leakage rate is less than 0.25%Vol/hr at -250 Pa.
Each chamber is equipped with an independent transfer chamber, with options for UV, pulsed xenon light, or atomization disinfection; CDCV(Cycle Development and Cycle Verification) services and comprehensive, reliable SOPs for material transfer are available.
Reserved VH2O2 interfaces for external vaporized hydrogen peroxide generators or optional built-in integrated VH2O2 systems.
Modular design with flexible configurations; the chamber and control cabinet can be integrated or placed separately, making it user-friendly for space-constrained scenarios.
Real-time monitoring of micro-environmental parameters including differential pressure,temperature, humidity, and ammonia concentration.
Built-in lighting system for operational convenience and ferret well-being.
| Model: LH7-1D-G6-BS | Model: LH7-1D-G8-BS |
| Overall Dimensions: L2700×W800×H1890mm | Overall Dimensions: L2040×W920×H2200mm |
| Chamber Dimensions: L1500×W800×H800mm | Chamber Dimensions: L1600×W920×H900mm |
| Number of Chambers: 2, arranged vertically | Number of Chambers: 1 |
| Operation Method: Single-sided operation, equipped with 3 sets of integrated sleeves on the operating face | Operation Method: Double-sided operation, equipped with 4 sets of integrated sleeves on each operating face |
| Sterile Path: Mechanical compression sealing, double-door interlocking; Optional disinfection methods: UV, pulsed xenon light, or atomized spray | Sterile Path: Mechanical compression sealing, double-door interlocking; Optional disinfection methods: UV, pulsed xenon light, or atomized spray |
| Waste Path: / | Waste Path: RTP Beta Container |
| Sealed Door: Tempered glass viewing window, mechanical sealing | Sealed Door: Tempered glass viewing window, mechanical sealing |
| Pressure Control: Adjustable from (-150) to 150 Pa; differential pressure between the interior and exterior of the chamber ≥ 50 Pa | Pressure Control: Adjustable from (-150) to 150 Pa; differential pressure between the interior and exterior of the chamber ≥ 50 Pa |
| Airflow Control: Air changes ≥ 20 times/h | Airflow Control: Air changes ≥ 20 times/h |
| Caging: Built-in ferret cage rack inside the chamber | Caging: Built-in ferret cage rack inside the chamber |
The relevant regulations and industry standards are as follows:
GB14925-2023 Laboratory Animal-Requirements for Environment and Housing Facilities
GB50346-2011 Architectural and Technical Code for Biosafety Laboratories
GB50447-2008 Architectural and Technical Code for Laboratory Animal Facilities
RBT010-2019 General Requirements for the Evaluation of Barrier and Isolation Devices for Laboratory Animals
RBT019-2019 Guidelines for Performance and Environmental Parameter Verification Procedures for Laboratory Animal Facilities
RBT199-2015 Technical Specifications for Biosafety Performance Evaluation of Laboratory Equipment