Opens in a new tab
  • HTY-CT2000

    TOC Analyzer

    High-temperature combustion + Non-dispersive infrared (NDIR)

    Tailin TOC Analyzer HTY-CT2000 uses high-temperature combustion oxidation. Samples are catalytically oxidized in an oxygen-rich environment at 680–1200°C. The generated CO₂ is carried into the non-dispersive infrared (NDIR) detector for measurement. TOC, TC, and IC can be measured separately.

    01
    Highly Sensitive TOC Detection
    02
    High-temperature Combustion Oxidation Tech
    03
    Linked Solid Combustion Device

    Model

    Features

    Solid Combustion Device: Can be equipped with a solid combustion device to measure solid samples such as soil and partially biodegradable plastics.

    Original Slider Isolation Design: Accurately isolates residual samples in the tubing, solving the industry challenge of residual water samples entering the combustion furnace and interfering with subsequent test data. It structurally avoids testing errors and ensures data accuracy.

    Digital Temperature Control Upgrade: Upgrades the traditional analog temperature control system to a PID integrated circuit digital temperature control module, offering faster response, higher temperature control precision, and more stable operation.

    Eight-Channel Rotary Valve: Direct acid addition and purging in the syringe: reduces IC redundant tubing by ~1/3, simplifies testing, and lowers contamination risks.

    Electronic Dual-Cooling Module and Membrane Dehydration Tech: Ensures the dehydration efficiency of the entire system to protect the NDIR.

    Can Be Connected with Automatic Sampler for Testing: Achieves intelligence and automation to improve the work efficiency.

     

    Specifications

    TOC Detection Range: (0–1000.000) ppm (without dilution)Detection Parameters: TC, TIC, TOC, NPOC, COD (by conversion only)
    Repeatability: ≤3% (JJG821-2005)Indication Error: TC: ±5%; IC: ±4% (JJG821-2005)
    Detection Limit: ≤0.1 mg/L (HJ501-2009)Quantification Limit: ≤0.5 mg/L (HJ501-2009)
    Max. Salt Content: ≤90 g/LSample Volume: Adjustable: 0.1–1.0 mL; recommended 0.3 mL
    Sample Method: Automatic eight-way valve injectionAmbient Temperature: (5–40)°C
    Relative Humidity: (10–85)% RHPower Supply: (220±10%)VAC, 50/60 Hz
    Rated Power: 1000 W (up to 1200 W during furnace heating)Automatic Dilution Ratio: 2–50×
    Gas Requirements: 99.999% pure oxygen / CO₂-free airGas Consumption: 120 mL/min
    Weight: 45kgDimensions: 550mm × 460mm × 525 mm

    Applications

    Can be used to test sewage and wastewater in the pharmaceutical industry.
    Can be used to test various types of wastewater, landfill leachate, and sludge in the environmental protection industry.
    Can be used to test various biodegradable materials and compost in the biodegradation industry.

    Pharmaceuticals

    food&beverage

    Food & beverage

    Municipal industries

    Send Inquiry