Opens in a new tab
  • STP Series

    Sterile Pass Box

    Turbulent flow

    GMP Grade A/B Sterile Transfer Chamber for Aseptic Manufacturing

    The STP Series Sterile Pass Box / Sterile Transfer Chamber achieve biological decontamination of the interior space, surfaces, and items (outer surfaces) through an integrated vaporized hydrogen peroxide (VH2O2) system. The integrated VH2O2 system is linked with the sterile pass box using SIEMENS PLC control. Communication between the sterile pass box and the VH2O2 system allows for the setting of relevant decontamination parameters, after which the system’s built-in program supports unattended automatic operation.

    01
    Turbulent flow
    02
    Comply with 21CFR Part 11, provides 4Q with on-site validation
    03
    Different size can be customized

    Model

    Features

    Using VH2O2 102/202 as a biological decontaminant, maintaining a low temperature and normal pressure during the decontamination process, which is efficient and environmentally friendly.

    The system provides real-time monitoring of the temperature, humidity, and pressure inside the chamber.

    Offering a complete 40-cycle program and on-site validation.

    During the biological decontamination and exhaust phases, all air entering the chamber is filtered through an H14-grade HEPA filter to prevent contamination of materials.

    The loading and unloading system features a double-door structure with pneumatic sealing, pneumatic locking, and an interlocking mechanism for the double doors during operation.

    The airflow inside the chamber is turbulent, facilitating gas diffusion when handling multi-layer loads.

    The loading and unloading doors are made of stainless steel frames and high-visibility tempered glass, sealed with pneumatic seals for excellent visibility.

    Specifications

    Airflow: TurbulenceCleanliness: GMP Grade A
    Biological Decontamination Cycle Time: 150 minutesBiological Decontamination Capability: Log reduction value≥6
    VH2O2 Residual Level: ≤1ppmHigh-Efficiency Filter Grade: Supply air HEPA (H14), Exhaust air HEPA (H14)
    Printing: Includes biological decontamination cycle parameter saving and printing functionsAlarm: High pressure/low pressure alarm, door open timeout alarm, door lock alarm, concentration/temperature alarm
    Exhaust Interface: 3-inch or 4-inch sanitary interfaceHydrogen Peroxide Gas Decontamination Cycle Interface: 1.5-inch sanitary interface
    Material of Inside Chamber: SUS316L stainless steel

    Applications

    It is widely used in microelectronics, biological laboratories, pharmaceutical factories, hospitals, and the food industry—any environment requiring air purification. It is suitable for transferring various clean and dry items in sterile production, including the outer packaging of packaging materials, instruments, raw materials, accessories, and environmental monitoring equipment that enter A and B critical areas.

    Pharmaceutical Industry

    Biotechnology Industry

    food&beverage

    Food and Beverage

    FAQ

    Send Inquiry

    Google reCaptcha: Invalid site key.