Opens in a new tab
  • HK Series

    Hydrogen Peroxide Gas Concentration Analyzer

    Used for detecting hydrogen peroxide sterilization applications

    The Hydrogen Peroxide Gas Concentration Analyzer employs TDLAS technology (Tunable Diode Laser Absorption Spectroscopy), utilizing the principle that specific wavelengths of light are absorbed by specific gases. This enables real-time monitoring of hydrogen peroxide and water vapor concentrations, while simultaneously calculating relative saturation and dew point temperature. By establishing a relationship with the D-value, the sterilization efficacy can be assessed through hydrogen peroxide and water vapor concentrations, allowing for the analysis and evaluation of VH2O2 and optimization of process parameters.

    01
    Utilizes TDLAS gas analysis method
    02
    Short response time
    03
    Strong adaptability to working conditions

    Model

    Hydrogen Peroxide Gas Concentration Analyzer
    In-situ HTY-HK100D
    In-situ HTY-HK200
    Bypass HTY-HK200D

    Features

    • Short Response Time: Real-time monitoring of gas concentration, with the fastest response time of 0.1s per data point.
    • Wide Detection Range: Hydrogen peroxide gas detection range (1-7000) ppm.
    • Strong Adaptability to Working Conditions: Can be used in high pressure, normal pressure, vacuum environment, high temperature up to 80℃, and takes up little space.
    • D Value Development: D value development models can be established based on the output dew point temperature and relative saturation.
    • High Accuracy: The linear error of hydrogen peroxide and water gas concentration is less than 5%.
    • Good Stability: Stable data and good repeatability ensure data reliability throughout the sterilization process.

    Specifications

    H2O2 Range: (1-7000) ppm (range adjustable according to actual applicationH2O2 Repeatability: ≤±5%
    H2O Range: (1500-30000) ppm (range adjustable according to actual application)H2O Repeatability: ≤±5%
    H2O2 Accuracy: ≤±5%H2O Accuracy: ≤±5%
    Ambient Pressure: (0-0.2) MPaCertification: CE
    Weight: HK200 (7.4kg), HK200D (10.74kg), HK100D (5.2kg)Detection Medium Temperature: (-20-80)°C
    Ambient Temperature & Humidity: (0-40)°C, (0-85)% RHDetection Time: 1s (fastest 0.1s)
    Signal Output: Supports RS485 digital signal output, 4-20mA analog signal output