Opens in a new tab
  • Hot Cells

    Hot Cells for R&D

    Used for R&D of radiopharmaceuticals, and other customized requirement

    In radiopharmaceutical research and development, the use of isolation and shielding technology minimizes exposure to operators and significantly reduces the risk of microbial contamination. High-contamination operations should be carried out in an isolated hot cell. Tailin R&D hot cell features a highly modular design to improve productivity and speed, with ergonomics that better meet the environmental requirements of radiopharmaceutical R&D.

    01
    Maximized Efficiency: Modular, ergonomic design accelerates radiopharmaceutical research from synthesis to testing.
    02
    Operator Safety First: Advanced shielding and isolation let researchers handle high-activity isotopes with confidence.
    03
    Seamless GMP Compliance: Intelligent control and real-time monitoring ensure traceable, reliable research data.

    Model

    Features

    Integrated system design: It integrates modules for synthesis, dispensing, material transfer, and waste handling, greatly enhancing operational convenience and safety in radiopharmaceutical R&D.

    Customized shielding: Designed based on radiation dose requirements, the hot cell effectively isolates and shields internal radiation, ensuring operator safety.

    Intelligent control system: Equipped with an advanced control and management system capable of real-time monitoring of parameters such as temperature, pressure, and radiation levels. It automatically adjusts operational status to maintain process stability and safety, and can generate GMP-compliant production reports and quality traceability record.

    Specifications

    Airflow: Turbulent flowCleanliness: GMP Grade C
    Chamber Tightness: At test pressure (2x working pressure), leakage rate <0.5% of chamber volume per hourShielding Performance: Dose rate s2.5 uSv/h at 30cm from any surface at the operator position Control System: Siemens PLC with 12" Siemens industrial panel PC
    Noise: <70 dB(A) during operationLighting: 500 lux inside the chamber, LED light source
    Power Supply: AC 220 V, 50 Hz*Customized according to user's requirements.

    Applications

    R&D of radiopharmaceuticals, and other customized requirement.

    Radiopharmaceutical R&D