In radiopharmaceutical research and development, the use of isolation and shielding technology minimizes exposure to operators and significantly reduces the risk of microbial contamination. High-contamination operations should be carried out in an isolated hot cell. Tailin R&D hot cell features a highly modular design to improve productivity and speed, with ergonomics that better meet the environmental requirements of radiopharmaceutical R&D.
Using the new cadmium zinc telluride (CdZnTe) material
Energy resolution of 2%
Operates at room temperature with a compact size.
Features
High-precision sample dispensing and high-precision camera imaging with a fixed light source enable accurate interpretation by AI algorithms.
It supports automatic sample spotting, plate development and detection, ensuring stable and consistent development results through fully automatic control.
It features a dynamic chromogenic method, automatic sampling and dilution, with mother liquor consumption as low as 5 µL and a detection time of 15–20 minutes.
Minimize radiation exposure of personnel during the testing process.
The amount of mother liquor used is no more than 1ml.
Specifications
Detectable Radionuclides: 18F, 68Ga, 177Lu, etc
Sample Requirement: ≤0.5ml
Test Items: pH, pH test paper, Endotoxin, Dynamic chromogenic method, Activity/Specific activity, Well-type ionization chamber
Liquid Distribution Function: Quality control equipment for liquid distribution
Surface Radiation Dose of Equipment: <2.5μSv/h
Lead Shielding Equivalent: 10-30mm
Weight: <1600kg
Table illuminance: 1000lx
Power Supply: 220V
Power: 2kW
Applications
Processing of radionuclides, such as 99mTc, 125I, 131I, 14C, 68Ga, 177Lu, 18F, 90Y, 225Ac, 89Sr and 89Zr