Opens in a new tab
  • GIT-WLAN03

    Glove Integrity Tester

    Used to leak test gloves installed on isolators or RABS

    The GIT-WLAN03 is used to leak test gloves installed on isolators or RABS. It leverages wireless communication and intelligent software technologies to achieve wireless device communication, paperless data management, and in-situ testing. It supports multi-device group control and data interconnectivity, and it automatically generates GMP-compliant electronic test reports, building a closed-loop data ecosystem that runs through the entire process.

    01
    AI algorithm deeply integrated with glove characteristics
    02
    Visualized testing process
    03
    The glove integrity management system can simultaneously monitor multiple glove integrity testers online

    Model

    Features

    Temperature Monitoring

    • The high-precision temperature sensor provides an accuracy of ±0.1 in detecting temperature differences during testing.

    Modular Design

    • The charging component module and internal battery pack can be easily disassembled and replaced.
    • The battery pack uses the internationally accepted standard 18650 battery.
    • Test port modules can be customized into different shapes and sizes according to customer need.

    Glove No. Code

    • RFID chip technology automatically identifies the glove number and includes it in the test results.

    Detection Accuracy

    • Leak rate detection limit can reach 100μm (under certain test parameters and test conditions).

    Convenience

    • Equipped with a charging trolley for continuous use, a fully charged machine can run for ≥8h.
    • Overall lightweight design, total weight ≤ 2.5kg, one person can easily hold and operate.

    Specifications

    Temperature Monitoring Accuracy: ±0.1℃Test Time: ≤6min
    Smallest Detectable Hole: 100μmContinuous Working Time: ≥4 hours
    Net Weight: ≤3KgsBattery: 12V Lithium battery
    Dimensions: Φ280x125mm*Glove flange interface can be customized

    Compliance

    ISO14644-7 Annex E.5

    “The pressure decay tests described are only a few of many tests that can be used for glove testing and are intended to illustrate glove leak test procedures.”

     

    EU GMP Annex 1 (2022) 4.21

    “For isolators, leak testing of the glove system should be performed using a methodology demonstrated to be suitable for the task and criticality. The testing should be performed at defined intervals. ”

    “For RABS… Gloves should be visually examined with each use, and integrity testing should be performed at periodic intervals.”

     

    FDA Guidance for Industry: Sterile Drug Products Produced by Aseptic Processing (2004) Appendix 1

    “…Physical integrity tests should also be performed routinely. A breach in glove integrity can be of serious consequence. …”

    Applications

    Reliable Glove Testing for Controlled Production Environments

     

    For isolators

    GIT-WLAN03 is designed for glove testing in isolator systems. Pressure decay testing helps ensure barrier integrity in sterile production environments and is suitable for routine inspections.

    For RABS

    In Restricted Access Barrier Systems (RABS), GIT-WLAN03 delivers reproducible glove integrity testing with configurable test parameters suitable for routine qualification and periodic checks.

    Pharmaceutical and Biotechnology Aseptic Production

    Food Industry Safety Control

    High-Risk Industrial Protection